Case   
Contact Us   
Your Position: Home > Case > Case 1

KCH602 portable ventilator

2016/1/19 11:10:02

No Pics.

Introduction
KCH602 Ventilator
The machine can be used in both adults and children, microcomputer control, LCD display, pneumatic control type ventilator, touch button set parameters, pressure beam monitor airway pressure, the machine electricity without electricity can be used, built-in lithium battery, if the battery is depleted of hand controlled breathing. Is the home, field, emergency vehicles, emergency room essential equipment.
1. Breathing mode:
Synchronous (A/C), mandatory (IPPV), manual
2, gas path performance:
Oxygen concentration: < 45%
3, ventilation performance:
A) respiratory rate: 6 ~ 60 times / minute, continuously adjustable.
B) the inspiratory time: 0.2 ~ 2S, continuously adjustable.
C) tidal gas quantity: 100 ~ 1000ml, continuously adjustable.
D) suction and call ratio: 1: 2, 1: 3: 4, 1: 1: 5: 1: 2: 2: 3: 1: 3: 5: continuous adjustable.
4, alarm performance:
A) the pressure limit alarm: 2 ~ 6kPa, continuous adjustable.
B) the pressure limit alarm: 0 ~ 2kPa, continuous adjustable.
C) battery under voltage alarm: sound and light alarm.
D): no less than 2min.
5, monitoring performance:
Tidal volume) a.
B) respiratory rate.
C) suction call ratio
D) synchronization time
6, machine performance:
A) power supply: AC 22V + + 50Hz + 1Hz
Built in lithium 7.2V power supply battery
B) host power: 25VA.
C) electrical safety: the relevant requirements of the GB9706.1-2007 II type B application parts for the part of the safety and general requirements of the medical electrical equipment.
7, the use of conditions:
A) environment temperature: +5 to +40.
B) relative humidity: 75% ~ 30%.
C) atmospheric pressure: 96 ~ 104kPa.
D) using gas source: oxygen is 0.25 ~ 0.3MPa.
8, storage and transportation:
The normal storage temperature range to - 18 DEG to 50 DEG C, relative humidity range from 15% to 95%, the atmospheric pressure range of 70kpa ~ 110 kPa, no corrosive gas and well ventilated place, transport process should be moisture-proof, handle with care, pay attention to the direction of the signs.

More Pics